Description:
Semi-automated 4-channel coagulation analyzer for fast and accurate determination of coagulation, chromogenic and immunturbidimetric assays. Compact in size, easy to operate, this analyzer is the ideal partner for the mid-size coagulation laboratory with slightly increased throughput.
The CoaDATA 4004 is the consistent continuing development of the existing turbodensitometric measuring principle. This unique measuring system combines a high-resolution photometer unit with the mechanical stirring of the test batch and ensures a precise and reliable determination of results over the entire coagulation spectrum.
Detection Principle
- LED Photometer (photo-mechanic)
- More insensitive to interferences from hemolysis, icteric and lipemic (HIL) samples
- Advanced coagulation diagnostics (coagulation, chromogenic and immunological methods) by selectable wavelengths (405nm & 750nm) for each measurement channel
- Magnetic stir bar for homogenizing the test approach and increase of sensitivity
- Large measuring range with higher accuracy and precision compared to only optical detecting principles
- Ambient light compensation
- Plasma Samples
- Micro volume: 150µl test volume
Measuring Unit (37,4°C controlled)
- 4 Measuring channels
- 4 Reagent positions (1x stirred)
- 16 Cuvette incubation positions
Features
- Internal thermal printer
- Open System, free editable reagent parameter
- 15 programmable test positions
- Automated calculation of seconds, INR, %, g/l, mg/dl, Ratio, mE, ng/dl
- 2 editable calibration-/reference curves (9 points)
- Patient-ID, numeric and alphanumeric (Barcode)
- Uni-directional HOST communication
- Multi-language user interface (DE, EN, ES, FR, PT, RU)
Dimensions: | 26 x 33 x 9 cm (LxWxH) |
Power supply: | 100V-240V, 47-63Hz, 260mA |
Output: | 24VDC / 3.75A |
Weight: | 3.9 kg |
Options: | External thermal printing device |
Interfaces: | ChipCARD® reading unit, RS232 (Printer / Barcode Scanner) USB-Device (HOST), SD-Card |
Coagulation Reagents
- PT (Seconds, %, INR)
- aPTT (Seconds, RATIO)
- Fibrinogen according to Clauss (mg/dl, g/l)
- Thrombin Time (Seconds)
- Batroxobin Time (Seconds)
Factor Deficient Reagents
- Extrinsic Factors II, V, VII und X (PT based)
- Intrinsic Factors VIII, IX, XI und XII (aPTT based)
Thrombosis Risk Reagents
- Protein C (Clotting Test)
- Protein S (Clotting Test)
- Lupus Anticoagulant
- Factor V Leiden
- APC-Resistance
Chromogenic Substrates
- Protein C (Chromogenic)
- Protein S (Chromogenic)
- Anti-Thrombin III
- Plasminogen
- α2-Antiplasmin
Immunturbidimetric Tests
- D-Dimer